UE9 Calibration App
A console C application (with VC6 source code) for recalibrating a UE9. The UE9 is shipped fully calibrated, but this application can be used to recalibrate at a later time if needed.
Requires UD driver V2.48+, a couple jumper wires, and a reference voltage meter with higher accuracy than the UE9. Uses either the first found UE9 over USB (make sure only 1 UE9 is connected), or the specified UE9 over Ethernet (make sure multiple UE9s do not have the same IP address).