A-3-3-3 T8 Signal Range [T-Series Datasheet]

The T8 analog inputs have channel-to-channel isolation allowing them to measure across their full range with common mode voltages up to the isolation spec from Appendix A-3-3-1.

Appendix A-3-3-1 has the abbreviated range values, which are also the values used when specifying the desired range. See the table below for the exact nominal ranges. For example, the T8 has a range called ±4.8V, specified for AIN2 by writing AIN2_RANGE=4.8, and the actual nominal range in that case is ±4.884V.

Range

Actual Range

Units

11

11.0[1]

±Volts

9.6

9.768

±Volts

4.8

4.884

±Volts

2.4

2.442

±Volts

1.2

1.221

±Volts

0.6

0.611

±Volts

0.3

0.305

±Volts

0.15

0.153

±Volts

0.075

0.076

±Volts

0.036

0.038

±Volts

0.015

0.019

±Volts

[1] The range is guaranteed to be at least 11 volts. The actual limit will vary between 11 and 11.3 V.