Production Status: NLA (no longer available). The Digit was discontinued in 2017.


Summary

It is possible to calibrate humidity to achieve higher accuracy. It is necessary to first leave the Digit in a chamber of known humidity for a few hours prior to calibration in software. There are multiple methods to calibrate hygrometers, please use online resources to find the a suitable method for your application.

After removing the Digit-TLH(s) from the humidity calibration chamber(s), plug them in and download the data. The calibration section in software is accessible after valid humidity data is downloaded. 

multi-data-download.png

After download is complete, open the calibration pane.

data-calibrate-h.png
multi-data-calibrate-h.png

This is the calibration section of Otero.  There are no parameters to calibrate Temperature or Light, so the calibration option will only appear if one or more Digit-TLH are connected to the computer.

multi-data-calibrate-h-main.png

The raw humidity value (Craw) is a measure of capacitance in hundreds of femto Farads.  To convert humidity from femto Farads into a %RH, use the following equation.

%RH = (Craw - CalOffsetH + ToffsetH)(Hslope) 

Read More: Opens a window with a few details about each parameter.

Restore Factory Settings:  Restores the factory calibration offset, slope, and temperature dependence of the Digit-TLH. 

Change any parameter on-the-fly to see the impact on downloaded data, or use the calculators to make it easy.

Calculate CalOffset: Enter a time into the time input box that corresponds with a known humidity value in the data set, then enter what the humidity value at that time should be, then click Calculate!. This requires time spent in 1 chamber of known humidity. The calibration offset will be automatically adjusted, and %RH will be the desired value.

Calculate Hslope: Enter two points into the time/value input boxes, then click Calculate! to automatically generate a new slope value. This requires time spent in 2 chambers of different known humidity. Note that after changing the slope, the CalOffset will need to be re-calculated: Just press Calculate again.

After the data looks correct, click Return, changes are auto-saved.*

*Starting with firmware v1.21, changes are auto-saved.  In previous versions of firmware, users had to click 'Apply' to save changes.